• The FIAlyzer 2000 is a multi-channel analyzers for simultaneous determination of multiple parameters. This instrument can accommodate up to 8 channels and offer the complete automation of wet chemical assays. Analysis of parameters like nitrate, nitrite, ammonia, phosphate, chloride and many others are routinely performed with the FIAlyzer series. By running these assays simultaneously, you achieve higher throughputs, reduce sample prep time, and save bench space.

    The FIAlyzer-2000 design is all about efficiency. With method lines tightly grouped, the system is easy to maintain and takes up little bench space.

    The LOV (Lab-on-Valve) platform, the centerpiece of this instrument, integrates manifold components (valve, tube fittings, confluence points) within a single transparent structure. It eliminates most of the tubing and connections typical for traditional flow injection manifolds.

    All of the instrument's reagent lines are located on the outside of the the instrument. This allows for simple and quick replacement of necessary lines. It also allows for easy inspection.

  • Analyzer

    Download Specifications

    Type: Two-Channel Flow Injection Analyzer 

    Enclosure material: Aluminum (powder-coated / anodized)

     

    Peristaltic Pump

    Type: 10-channel peristaltic, common pressure plate

    Materials: Aluminum, stainless steel

    Tubing type: 2-stop elastic tubing (Tygon, Pharmed etc)

    Flow rate: 0.5-5.0 mL/min per channel (1.02 mm ID tubing) 0.3-2.8 mL/min per channel (0.76 mm ID tubing)

     

    Valve

    Type: 12-port; Two Position InjectionValve

    Wetted materials: Plexiglas, Valcon E2 (proprietary reinforced PTFE composite)

    Port type: Flat-bottom 1⁄4-28 Relay output Voltage: 24 V Current: Max 4mA

     

    Dimensions

    Height: 11.5 in (29 cm)

    Width: 7 in (18 cm)

    Depth: 9.5 in (24 cm)

    Weight: 18 lb (8 kg)

     

    Communication

    Input: USB

    Output: none

    Control options: FIAsoft software

     

     

    Power requirements

    Voltage: 24 Vdc

    Current: Max. 2.5 A 

  • FIAlyzer-2000 Manual

    FIAlab Product Catalog

Product Highlights

The FIAlyzer-2000 is a robust instrument with a lot to offer.

Fast Throughput

The FIAlyzer-1000 is the fastest flow injection analyzer on the market. Analyze up to 250 samples per hour with this instrument.

Small footprint

Save room in your laboratory. This instrument requires less than one square-foot of bench space.

Spectrometer Capabilities

Our analyzers utilize spectrometers to perform their detection. A spectrometer will broaden your sample concentration range, reduce noise, and collect more data.

Free Training and Installation

Every instrument comes with free training and on-site installation. Our technicians will teach you the in-and-outs, run your samples, and ensure that the instrument accomplishes your analytical goals.

Software

Software is just as important as hardware. In order to provide complete analytical solutions, we understand that we have to deliver quality hardware and polished software. Our instruments operate on the newest, most intuitive FIA software ever. This program will change the way you use flow injection. Click the link to find out how.

Typical Setup

...

Analyzer

Choose the analyzer that suits your application best. Is it single channel, multi-channel or a combination?
...

Computer

We supply computers at cost. We will install all relevant software and hardware. Or, use your own. Up to you!
...

Autosampler

We can supply an autosampler at retail value, or you can use one you already own. FIAsoft is compatible with all standard autosamplers.
...

Service Contract

Consider purchasing a Preventative Maintenance Plan with FIAlab. With it, you will receive unlimited consultation, an extended warranty, and annual maintenance visits.

Methods

Ammonia

Gas diffusion and salicylate methods. EPA recommended.

Nitrate

Cadmium reduction. ASTM 4500-NO3- I

Ortho-Phosphate

Automated. ASTM 4500-P G-199

TKN

Manual digestion and distillation followed by any salicylate. Standard Method 4500- Norg D-

Combine any two methods to run simultaneously…

Request a Quote