Tungsten Lightsource 3 of 7 01

Text:

The DH-2000 light source (PN 17002) offers stable, continuous output from 215-2500 nm. Using a combination of deuterium and halogen lamps, the DH-2000 is ideal for measuring a sample that has multiple features in different spectral regions or for analyzing a variety of different samples in your lab. All models of the DH-2000 family include easy-to-replace bulbs, keeping your measurements going with fast lamp changes.

For those that need light into the deep UV, the DH-2000-FHS-DUV-TTL and DH-2000-S-DUV-TTL models include a special deep-UV deuterium lamp providing output from 190-2500 nm. Both models come with a shutter to limit the exposure of samples to the UV light.

The DH-2000-FHS-DUV-TTL comes with a filter holder for filters up to 25 mm square or 25 mm round in diameter and 4 mm in thickness, plus a shutter controlled via TTL signal or switch up to 5 Hz.

Specifications

Engineering Specifications

DH-2000-DUV Series

Sources:

Deep-UV Deuterium & Tungsten Halogen

Wavelength range:

190-2500 nm

Nominal bulb power:

26 W (deuterium)
20 W (tungsten halogen)

Typical output power:

See chart below

Warm-up time:*

25 minutes

Source lifetime:

1,000 hours

Stability of light source output:

≤0.1%/hour @ 254 nm (deuterium)

≤0.1%/hour @ 700 nm (tungsten halogen)

Drift:

≤0.1%/hour @ 254 nm (deuterium)

≤0.1%/hour @ 700 nm (tungsten halogen)

Shutter:

DH2000-S-DUV & DH2000-S-DUV-TTL only

Remote shutter control:

DH2000-S-DUV-TTL only

Integrated filter holder:

DH2000-FHS-DUV only

Operating temperature:

5 °C – 35 °C

Operating humidity:

5-95% without condensation at 40 °C

Power requirements:

85-264 V, 50/60 Hz

Power consumption:

Approximately 78VA

Dimensions (W x H x L):

15 x 13.5 x 28.5 cm

Weight:

5 kg

Safety & regulatory:

CE; ROHS, WEEE

Replacement bulbs:

DH-2000-DUV-B (deuterium)

DH-2000-BH (tungsten halogen)

*  Warm-up time @23 °C ambient, free airflow, no vibrations

Output Power

The table below reports typical output power coupled into a 600 μm UV-VIS fiber and measured with an optical power meter integrated from 200-1100 nm.

Model

Deuterium Bulb Only

Tungsten Bulb Only

DH-2000-FHS

217 μW

295 μW

DH-2000-DUV

585 μW

  1. μW

Shutter/Trigger Options

Available only for the listed models:

Engineering Specifications

DH2000-S-DUV-TTL

Trigger/shutter input signal:

TTL; Up to 2.5 Hz maximum

Trigger/shutter connection:

SUB-D-15 pin

Independent switching of deuterium and the halogen bulbs

Yes, via TTL signal

DH 2000 fam spectra v