Sulfate  

FIAlab standard method for sulfate assay using the FIAlab-2500/2600/2700 system.
Assay Typical Throughput Concentration Range Notes
Sulfate (Mid-high) 120 samples/hour 1 to 50 mg Cl-/L 1 cm flowcell
Sulfate (Low) 60 samples/hour

0.1 to 5 mg Cl-/L

10 cm flowcell

 


Principle:

Sulfate in the sample is precipitated with acidified barium chloride.  This precipitate scatters light producing a signal proportional to sulfate concentration.   The following method should work for concentrations from 10 ppm to 200 ppm.  Please inquire for suggestions on how to measure lower or higher concentrations than these ranges.


Comments:

For most Sulfate assays make the sample loop from 12 inches of 0.03" ID tubing.   Use a 10 cm flowcell for concentrations of 10 ppm to 200 ppm.  Use a 1 cm flowcell for concentrations consistently greater than 100 ppm.   Recommended monitored wavelength is 500 nm, though any wavelength from 400 to 800 nm will work.  Do not use a reference wavelength.

Sulfur can be added to the reagent to “seed” the precipitate.  This is necessary for the lower concentrations.


Interferences:

Color and suspended particulates will interfere. Sulfites and sulfides may oxidize and then precipitate as barium sulfate. Large quantities of organic matter interfere.

Reagents: 

Carrier: Matrix Match
Matrix Match.
  If samples are water then use water as carrier.  If soil extract, then (typically) add KCL in proportion with the extracted samples, e.g., 12 g KCL per liter of water.   The matrix matching isn’t critical if samples have greater than 100 ppm Sulfate concentration.

Reagent 1: Barium Chloride + HCL + Sulfur (for one liter)

30.0 grams BACl  Dihydride  

10 ml  Hydrochloric acid 

0.004 grams Sulfur of DI Water (this is not needed for sample sulfate conc. > 10 ppm)

1-Liter  DI Water.    Stir until dissolved.

Reagent 2: No Reagent 2 – plug this channel port


Standards:

100ml ICSUL-100  (Sulfate standard)
Source: 727-524-7732 -  www.exaxol.com 

 