Special Hardware:
FIAlab - IL-UV
In-line Digestion Unit for
In-line digestion of total
nitrogen, total phosphorous, total
cyanide. Includes secondary two channel
peristaltic pump.
Programming:
Please contact FIAlab
for details on the programming/timing of
the sample digestion period within the
IL-UV coil.
Reagents:
Carrier:
The carrier should be matrix matched to
the sample, i.e., salt content for
seawater samples, and KCL content for
soil extracts.
Persulfate
Digestion Solution:
7.6 g of sodium
tetraborate decahydrate
16.0 g of potassium
persulfate
Dissolve in 1 L DI
water
Reagent 1: 1.6 M Ammonium Chloride
Buffer (for nitrate only).
Note - For
nitrite,
plug this R1 port with an included
Teflon Plug
500 ml TOTAL: Balance with Degassed DI water
43 grams
Ammonium chloride
0.50 grams
disodium ethylenediamine tetraacetic
acid (EDTA) [372.24 FW]
(only if metals may be present)
~1 grams Sodium Hydroxide Pellets
as needed
Place 400 cc of water into a volumetric
flask and add the ammonium chloride and
EDTA.
Mix until dissolved.
Adjust the pH of this solution to
~8.5 using sodium hydroxide pellets.
Add water to fill flask to 500mL.
CAUTION:
Ammonium chloride sometimes may
contain significant amounts of nitrate.
Reagent 2: Colorimetric Sulfanilamide
Solution.
20
grams
Sulfanilamide [172.21 FW]
50 ml 85% Conc. Phosphoric Acid
(H3PO4)
0.50 grams
N-1-naphthylethylenediamine
dihydrochloride [259.18 FW]
500 ml TOTAL:
Balance with Degassed DI
water
Mix well and store in dark
bottle.
Cadmium Column:
Cadmium column can be prepared/packed by
user, or purchased ready for use.
For nitrite assays, use a
bridge of 0.03" ID tubing (~4 inches) in
place of the column.
1)
CC120
Cadmium Column for nitrate reduction
column
(or)
2)
Prepare your own Cadmium
Standards:
100ml
ICNAT-100 (nitrate standard)
100ml
ICNIT-100 (nitrite standard)
Source: 727-524-7732 -
www.exaxol.com
|