FI - Lab on Valve®


Email: fialab@flowinjection.com 
Voice 425-376-0450 or 1-800-963-1101, Fax 425-376-0451

DOUBLE CHANNEL    
The LOV modules are standard components of FIAlab 2500  and FIAlab 2600 instruments.
The advantage of LOV design is transparency of the entire flow path, that allows the user
visual inspection of all manifold components, including injector valve and flow trough cell,
at any time, without interruption of ongoing automated assays.

Single Channel FI-LOV

Sample Injection by Two Position Valve

The Flow Injection technique employs a two position valve for injection of a well defined sample volume into the carrier stream containing regents. The sample volume is defined by the volume of an external sample loop – selectable by the user. The sample is aspirated into the loop in the LOAD position and inserted into the carrier steam after the rotor of the two position valve has moved into the INJECT position.

LOAD
INJECT
Single Channel FI lab-on-valve® System


The FI –LOV manifold comprises a two position injection valve, furnished with an external loop. The valve is integrated with a manifold is equipped with T joints where the carrier stream merges with reagent streams. The structure has threaded outlets that accommodate fittings and  mixing coils along with a flow cell for UV-VIS spectrophotometry, connected to light source and detector by means of optical fibers. The LOV  manifold is available either in Perspex as shown here, or in ULTEM.

Single Channel FI-LOV Manifold

For further details see FIAlab 2500 and Flow Injection page

Dual Channel FI-LOV

This unique valve configuration has been designed to  save cost and bench space, and to accelerate sampling rate (sample has less distance to travel to fill both sample loops), The result is a compact instrument suited for laboratories performing routine assays at an yet unprecedented
productivity and time saving.
For further details see FIAlab 2600

For information  on applications see Agricutural and
Environmental Assays page.

Dual Channel FI-LOV in ULTEM


Specifications

FI-LOV fabricated from Plexiglas; fully transparent, wetted surfaces PEEK, Teflon and Plexiglas. Compatible with water  solutions of reagents, acids and bases
(up to 0,5M) and ethanol or methanol /water  solutions up to 10%. w/w.
FI-LOV fabricated from ULTEM: partially transparent, wetted surfaces PEEK, Teflon, ULTEM.

Partial Chemical Compatibility list for ULTEM:

Acetic Acid, Acetone Alconox Acetone, Alconox , Carbon Tetrachloride, Cellosolve, Chromic Acid, Citric Acid, Clorox, Cyclohexane, Cyclohexylamine, Ethanol, Ethyl Acetate, Ethyl Ether, Formic Acid, Gasoline, Hexane, Hydrochloric Acid, Isopropanol,  Kerosene, Lestoll 1, Methanol, Methylethyl Ketone (MEK), Naphtha, Nitric Acid, Phenol, Phosphoric Acid, Potassium Carbonate, Potassium Hydroxide, Propylene Glycol, Skydrol, Sodium Hydroxide, Sulfuric Acid, Tetrachloroethylene, Tin Chloride, Toluene, Trichloroethane (1,1,2-), Triethylphosphate, Xylene, Zinc Chloride    

 
 
 
Examples of Analyte Combinations
Throughput: Assay Results
Nitrate and Nitrite
240
Nitrate and Ammonia
120
Nitrate and Chloride
240
Nitrate and Sulfate
120
Ortho Phosphate and Nitrate
240
Ortho Phosphate and Ammonia
120
Bray Phosphate and Potassium
240